8-4
In the “General Setup” tab, you can set high alarm volume. High alarm volume can be set
equal to or louder than the alarm volume. The choices are “alarm volume +0”, “alarm
volume+1”and “alarm volume+2”. When a high level alarm occurs, the alarm will sound with
the specified high alarm volume. For example, if the high alarm volume is set to “alarm
volume+1”, and the current level of alarm volume is 5, then the high alarm volume level is 6.
NOTE
z If high alarm volume level is greater than 10, the monitor will sound it in the
volume of 10.
8.4 Alarm Setup
In either the “Parameter Setup” or “Alarm Setup” tab sheet, you can define alarm switch,
alarm high/low limits and alarm priority for each parameter.
8.4.1 Parameter Setup Tab Sheet
Click in a parameter window. The border of the window will appear highlighted when the
window is selected. Clicking again will enter a tab sheet.
Take the ECG module as an example; you can set the HR, ST Analysis and Arrhythmia
Analysis, such as alarm limit and alarm priority.
NOTE
z The settings of the parameter alarm switch, limit and level are controlled
bidirectionally. For example, if the alarm switch of a parameter is set to “Off”,
the alarm switch of this parameter will change to “Off” on the monitor.
z When you change the alarm limits of the monitor via the CMS, make sure that
the alarm limits settings are appropriate for your patient. Setting alarm limits to
extreme values may cause the alarm system to become ineffective.
Default: Click this button to restore CMS-related settings to factory default.
Alarm Setup: Click this button to enter the Alarm Setup tab sheet. Refer to Section 8.4.2
Alarm Setup Tab Sheet Alarm Setup Tab Sheet for details.
For telemetry transmitter, the parameter setup tab sheet is different. For details, refer to the
manual of telemetry monitoring system.