12-7
Select Stored Waveforms
In the “Alarm” tab sheet, you can select the relevant waves stored when alarms occurred to
the specific parameter.
1. Click the “Alarm” tab sheet.
2. Select a parameter from the “Alarm Numeric” list. Then select the relevant wave stored
from the “Related Waveform” when the alarm occurred.
3. Click “Save” to save the selected wave.
4. Click “Default” to restore all parameters to the default configuration.
NOTE
z No matter how many waveforms are selected from the Related Waveform list,
the system can at most save 3 waveforms and record 2 waveforms for one
parameter when an alarm occurs.
z Your selections are only effective for later alarms. For already existed alarms, it
is unnecessary to re-select waveforms.
12.2.4 Trend Group
You can set the parameters in “Trend Review” by group.
The trend group tab sheet contains system default trend groups list and custom trend groups
list, and the list of corresponding trend group member parameters as well. The system
provides nine (9) default trend groups, which are “All”, “Standard”, “ECG”, “GAS”,
“Hemodynamic”,“RM” and “BIS”, “Ventilator” and “Anesthesia”. You cannot edit the
default trend groups.
Create trend group: click “New Group” and enter the name of custom trend group in the
pop-up dialog box.
Edit trend group: after selecting custom trend group, you can select/deselect the member
parameters or adjust the order of member parameters in the list.
Save trend group: after creating or editing trend group, click “Save Group” to save the
new or modified trend group.
Delete trend group: you can delete any custom trend group from the trend group list.
Rename trend group: click the custom trend group to be renamed, enter the new name of
the trend group in the pop-up dialog box.