2-2
2.3 Contraindications
None.
2.4 Functions
The CMS comprises powerful system software and high-performance computer components.
It constructs a monitoring network by connecting monitors and telemetry transmitters. By
collecting, processing, analyzing and outputting the information coming from monitors and
telemetry transmitters, the CMS can achieve centralized monitoring over multiple patients so
as to greatly promote the efficiency and quality of the monitoring work.
The CMS:
Is capable of connecting up to 32 monitors that supports the CMS or CMS+ network
protocol.
Is capable of connecting up to 32 telemetry transmitters.
Supports any bed’s (1-16 bed) frequency setup by each telemetry monitoring system.
Supports multi-screen display mode, including up to four local displays and multiple
remote displays; supports remote large screen display.
Is capable of displaying information from 16 monitors in the single-screen mode and 32
monitors in the dual- or multi-screen mode.
Allows you to view a single patient emphatically.
Supports the networking between multiple CMSs and the “Remote CMS”.
Allows you to review up to 240 hours of trend data for each online patient.
Allows you to review up to 720 events for each online patient.
Allows you to review a 4-hour dynamic short trend for each online patient.
Allows you to review up to 240 hours of ST segments for each online patient.
Allows you to review up to 240 hours of waveforms for each online patient.
Allows you to review up to 240 hours of ECG compressed waveforms for each online
patient.
Allows you to review up to 240 hours of NIBP measurements for each online patient.
Is capable of storing discharged patients’ full disclosure waveforms, parameters, alarms
and so on ; allows you to search and review the data of up to 20,000 discharged patients.
Provides the patient information management function.
Provides audible and visual alarms.
Provides the functions of drug calculations, titration table calculations, hemodynamics
calculations, oxygenation calculations, ventilation calculations and renal calculations.
Provides the functions to record, print and save data.