3-6
3.4 Shutting down the System
It is important to shut down the system properly. Follow this simple procedure to properly
shut down your system. This prevents inadvertent errors from occurring during system shut
down.
1. Click “System Setup” button.
2. Select “Shutdown” from “General Setup”.
3. The system will check if any patient is being monitored.
If no patients are being monitored, directly enter the next step.
If there are still some patients being monitored, the message box “× patients are
being monitored. Are you sure you want to shut down?” will be displayed. You can
either select “Yes” to enter the next step, or select “No” to close the message box and
not continue the shutdown process.
4. A message box asking whether to shut down the system will be displayed. Select
“Yes” from the message box.
If no shutdown password is required, the CMS will quit with the operating system shut
down and the host powered off.
If a shutdown password is required, a dialog box will pop up. You should enter the
required password and select “OK”. The CMS will quit with the operating system shut
down and the host powered off
5. If an UPS is used, turn it off.
6. Disconnect the power cord of each device from the power outlet.
CAUTION
z Hospitals without a stable power source should use a UPS to provide power to
the CMS. The UPS must not be turned off. When there is a power failure, the
system should be shut down by following the specified shutdown procedure
before the UPS is exhausted. When the system has a sudden power failure,
system failure may occur and consequently the system will not work correctly
next time or even have a serious result.