A-15
Resolution 1 bpm
Pressure transducer
Excitement voltage 5 VDC, ±2%
Sensitivity 5 uV/V/mmHg
Impedance range 300 to 3000
Volume displacement
(ABBOTT)
0.04 mm
3
/100 mmHg
A.14 CO
2
Specifications
Measurement technique Infrared absorption technique
Displayed parameter EtCO2, FiCO2, Respiration Rate
CO2 function
Meet the requirements of EN ISO21647/ISO 21647 and
ISO9918.
A.14.1 Mindray DS CO
2
Specifications
CO2 measurement range 0 to 99mmHg
Precision*
0 to 40 mmHg:
41 to 76 mmHg:
77 to 99 mmHg:
±2 mmHg
±5%
±10%
Resolution 1 mmHg
Drift meet the requirement of accurancy in 6 hours
Sample flow rate 70, 100 ml/min
Precision of deflation rate ±15% or 15 ml/min, whichever is great
Start-up time of CO2
module
< 1min, the module enters the warming up status after the startup.
One minute later, it enters the ready-to-measure status.
AwRR measurement range 0 to 120 BrPM
Precision
0 to 70 BrPM:
> 70 BrPM:
±2 BrPM
±5 BrPM
Response time
When measured with a neonatal watertrap and a 2.5 m-long
neonatal sampling line:
<3.5 s @ 100 ml/min
<4 s @ 70 ml/min
When measured with an adult watertrap and a 2.5 m-long adult
sampling line:
<5.5 s @ 100 ml/min
<7 s @ 70 ml/min
Delay time
When measured with a neonatal watertrap and a 2.5m-long
neonatal sampling line: