16-7
O2 COMPEN The options include: OFF, 30%, 40%, 50%, 60%, 70%,
80%, 90%, and 100%.
When the O2 concentration exceeds 60% and it is not being monitored, turn on this switch.
SWEEP Used to select the speed to scan the screen waveforms.
Options: 6.25mm/s and 12.5mm/s.
WORK MODE Options: MEASURE and STANDBY.
To monitor the anesthetic gas, select the MEASURE option. Otherwise, select the STANDBY
option.
Alarm Setup Menu
Select ALARM SETUP >> in AG SETUP menu. The following menu appears.
Figure 16-5 Alarm Setup Menu
In the ALARM SETUP menu-1, you can set the following items.
ALM Alarm on/off status
ON: When the AG has an alarm, the system will give alarm
indications and save the alarm information.
OFF: The system will not give alarm indications when an
AG alarm occurs. Instead it will display
on the right of
the GAS label in the parameter window.
ALM LEV Options: HIGH and MED.
ALM REC ON: The recorder outputs the alarm information in case an
AG alarm occurs.
OFF: The recorder does not output the alarm information in
case an AG alarm occurs.
EtCO2 ALM HI Set the upper limit of EtCO2 which triggers the alarm