10-30
10.7 12-Lead
10.7.1 12-Lead Monitoring Display
You can enter the 12-lead monitoring display if your monitor is configured with the 12-lead
monitoring. To enter the ECG 12-Lead monitoring screen:
1. Place the 10 electrodes on the patient.
2. Select the ECG label to enter the ECG SETUP menu.
3. Set LEAD TYPE to 12 LEADS.
4. Set the ECG DISPLAY to 12-LEAD DISPLAY.
Figure 10-21 12-Lead Display
A total of 12 ECG waves and 1 rhythm wave will be displayed on the screen. The ST numerics
are displayed in three groups:
ANT (anterior): V1, V2, V3, V4
INF(inferior): II, III, aVF, (aVR)
LAT (lateral): I, aVL, V5, V6
Although aVR is displayed in the INF group, it is not an inferior lead.
Rh
y
thm l
a
la
l