Mindray TRIO Provozní pokyny Strana 94

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 170
  • Tabulka s obsahem
  • ŘEŠENÍ PROBLÉMŮ
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 93
Trio™ Operating Instructions 0070-10-0666-01 2 - 65
Operations Alarms
COMMON ALARM SETUP Menu Selections
FIGURE 2-47 Common Alarm Setup Menu
ALM SEL
(Alarm Selection)
Choose COMMON ALARM SETUP.
ALARM VOL Volume of the audio alert. Selections are: LOW, MED, and HIGH.
ALM PRINT TIME Length of the printed waveform. Selections are: 8s, 16s, or 32s.
ALM MUTE TIME Length of time for the alarm to be silenced. Selections are:1min,
2min, 5min, 10min, and PERMANENT.
When ALM MUTE TIME is set to PERMANENT and Alarm Mute
is active, the message “Alarms Permanently Muted” displays at a
constant rate in the technical alarms area.
If another technical alarm occurs at the same time, both
messages will display alternately.
•If the ALM MUTE TIME is changed to something other than
PERMANENT, the following conditions will exist:
The elapsed time is defined as the time from which ALM
MUTE TIME had been set to PERMANENT to the time at
which it was changed to something other than
PERMANENT.
If the elapsed time exceeds the newly selected ALM MUTE
TIME, then the Alarm Mute will immediately deactivate.
Example: Elapsed time = 5min, New ALM MUTE TIME =
2min, Alarm Mute immediately deactivates.
If the elapsed time is less than the newly selected ALM
MUTE TIME, then the Alarm Mute will continue to be active
for the remainder of the newly selected time period.
Example: Elapsed time = 2min, New ALM MUTE TIME =
10min, Alarm Mute continues to be active for 8 minutes and
then deactivates.
Zobrazit stránku 93
1 2 ... 89 90 91 92 93 94 95 96 97 98 99 ... 169 170

Komentáře k této Příručce

Žádné komentáře