8-5
8.4 Setting Initial Cuff Inflation Pressure, NIBP Reset,
Calibration and Test for Air Leakage
8.4.1 Setting Initial Cuff Inflation Pressure
In SYSTEM SETUP menu, select MAINTAINNIBP TOOLSInitial Cuff Inflation
Pressure, and then you can set the initial cuff inflation pressure. The initial cuff inflation
pressures are listed in the following:
Adult80100120140160180200220240 mmHgDefault is 160 mmHg
Pediatric80100120140160180200 mmHgDefault is 140 mmHg
Neonate6080100120mmHgDefault is 100 mmHg
NOTE
The cuff will be inflated with the default initial cuff inflation pressure at each
time you set the measurement mode as MANUAL.
8.4.2 Resetting NIBP
If the blood pressure pump works incorrectly but the monitor does not alarm for it, you can
check the pump by resetting it. To reset the pump, select SYSTEM SETUPMAINTAIN
NIBP TOOLSNIBP RESET.